Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Condumax CLS15D is a reliable, digital conductivity sensor for pure and ultrapure water applications

Condumax CLS15D

CLS15D
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Conductive

  • Application

    Measurement in pure and ultrapure water range, monitoring of ion exchangers, reverse osmosis, distillation, chip cleaning

  • Characteristic

    Analog or digital 2-electrode conductivity sensors

  • Measurement range

    k=0,01: 0.04 to 20 µS/cm
    k=0,1: 0.10 to 200 µS/cm

  • Measuring principle

    Conductive conductivity cell with polished stainless steel electrodes

  • Design

    2-electrode conductivity cell with coaxially arranged electrodes, electropolished

  • Material

    Electrode: stainless steel 1.4435
    Electrode shaft: PES

  • Dimension

    Electrode diameter: 16 mm (0.63 inch)
    Electrode length: appr. 55 mm (2.17 inch)

  • Process temperature

    Threaded with fixed cable:
    -20 to 100 °C (-4 to 212 °F)
    Threaded with plug-in head:
    -20 to 120 °C (-4 to 248 °F)
    Sterilization: max. 140 °C (284 °F) for 30 minutes

  • Process pressure

    13 bar at 20 °C (188 psi at 68 °F)
    1 bar at 120 °C (14 psi at 248 °F)

  • Temperature sensor

    NTC 30K

  • Ex certification

    ATEX, NEPSI, FM, CSA

  • Connection

    Process: 1/2" and 3/4" NPT, Clamp 1 1/2"
    Cable: Memosens connector

  • Ingress protection

    IP68

  • Additional certifications

    Calibration certification of the cell constante, material certification 3.1