Hãy thử tìm kiếm theo từ khóa, đơn hàng hoặc mã sản phẩm hoặc số sê-ri, ví dụ: “CM442” hoặc “Thông tin kỹ thuật”
Nhập ít nhất 2 ký tự để bắt đầu tìm kiếm.
  • Lịch sử
Liquiline System CA80NO - Nitrite analyzer for drinking water, mineral water, wastewater

Nitrite analyzer
Liquiline System CA80NO

Colorimetric system for monitoring of drinking water, mineral water or raw water for food production

CA80NO
Please Wait (spinning wheel)
Đang tải giá
Giá tạm thời không có sẵn
Price on request
from CHF ??.-
Lọc phép đo / Tác vụ ứng dụng
Nhiệm vụ đo / Ứng dụng

Định cấu hình để tải xuống các bản vẽ CAD của sản phẩm này

  • Measuring principle

    Colorimetric

  • Characteristic

    Process analyzer for Nitrite in potable water

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 to 40 °C (39 to 104 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)
    Outdoor version: - 20 to 40 °C (-4 to 104 °F)

  • Process pressure

    At atmospheric pressure, < 0.2 bar absolute

  • Sample flow rate

    Min. 5 ml/min (0.17 fl.oz/min)

  • Consistency of the sample

    Suspended solids content
    Turbitity < 50 NTU, aqueous, homogenized

  • Specials

    Easy upgrade to measuring station with up to four digital Memosens sensors
    Automatic calibration and cleaning
    User-configurable measuring, cleaning and calibration intervals
    Optional cooling module for standard solution
    2 channel version optional
    User-definable measuring ranges
    Modular design for easily extensible functionality

  • Application

    Limit value monitoring of nitrite for potable water and mineral water applications

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 or 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 to 20 mA
    Modbus RS485/TCP (optional)
    Ethernet (optional)
    Alarmrelay

  • Input

    1 or 2 measuring channel
    1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    10 to 500 µg/l NO2-N
    0.2 to 3.0 mg/l NO2-N
    0.1 to 1.0 mg/l NO2-N
    0.1 to 1.0 mg/l with dilution function to maximum 5 to 50 mg/l NO2-N

  • Consumables

    Reagents and standard solutions CY80NO as well as cleaner CY800 are necessary for the operation
    Regular maintenance is done with the parts of the maintenance kit CAV800

Chào mừng đến với Endress+Hauser!

Vui lòng chọn thị trường / khu vực của bạn