Tổng quan về thông số kỹ thuật
-
Process temperature
4 to 40 °C (39 to 104 °F)
-
Process pressure
1.5 to 4.0 bar (21.76 to 58.01 psi)
For optional automatic cleaning:
2.0 to 5.0 bar (29.0 to 72.5 psi); but at least 0.5 bar (7.3 psi) > process pressure
Phạm vi ứng dụng
Liquiline System CAT810 provides filtrate for continuous online monitoring of municipal and industrial wastewater and industrial processes. Its self-cleaning crossflow filter avoids filter blocking, thus increasing your process safety. Thanks to a minimized dead volume, it reflects changes in your process promptly and shortens the response time of your downstream measuring devices.
Liquiline System CAT810 is designed for:
Outlets of municipal and industrial wastewater treatment plants
Collection of aqueous samples from pressurized pipes in industrial processes
Lợi ích
Long maintenance intervals due to automatic back-flush function
Save costs through tool-free maintenance and easy replacement of filter elements
Control via the user-friendly Liquiline System analyzer for perfect synchronization of the measuring point or timer-controlled operation
Various mounting options for optimum fit to the application
Short response time of downstream measuring devices thanks to low dead volume
Sản phẩm cũ
-
Sản phẩm chuyên dụng
- Dành cho các ứng dụng đòi hỏi cao
Công nghệ vượt trội
Đơn giản
Có thể thay đổi
Lựa chọn FLEX | Công nghệ vượt trội | Đơn giản |
---|---|---|
Lựa chọn Fundamental Đáp ứng nhu cầu đo lường cơ bản |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Lean Xử lý dễ dàng các quy trình cốt lõi |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Extended Tối ưu hóa quy trình bằng công nghệ tiên tiến |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Xpert Làm chủ các ứng dụng khó khăn nhất |
Công nghệ vượt trội
|
Đơn giản
Có thể thay đổi |
-
Measuring principle
Colorimetric
-
Characteristic
Automatic sample preparation system for supplying process measuring devices with filtered samples from pressurized pipes
-
Size
Panel version:
870 x 350 x 177 mm
34.25 x 13.78 x 6.97 inch
Prepared for stand housing CA80:
828 x 327 x 327 mm
32.6 x 12.87 x 12.87 inch
Basic version:
811 x 303 x 131 mm
31.93 x 11.93 x 5.16 inch -
Design
Back-wash filter system with various mounting options:
Panel version, Prepared for stand housing of CA80, Basic version - filter unit -
Process temperature
4 to 40 °C (39 to 104 °F)
-
Ambient temperature
4 to 40 °C (39 to 104 °F)
-
Process pressure
1.5 to 4.0 bar (21.76 to 58.01 psi)
For optional automatic cleaning:
2.0 to 5.0 bar (29.0 to 72.5 psi); but at least 0.5 bar (7.3 psi) > process pressure -
Specials
Optional back-flush with pressurizd air or water
Controling and parametrition via Lquiline System CA80
Stand-alone version with time-control available -
Application
Sampling from pressurized pipes
Preliminary and secondary clarification of municipal and industrial wastewater treatment plants
Sampling of aqueous media from industrial processes -
Power supply
100 to 120/200 to 240 V AC ±10%, 50/60 Hz
Chọn theo đề xuất của chúng tôi
-
-- selected coding-- selected label99.999,00<%label%>
-
Bạn muốn có nhiều lựa chọn hơn? Vui lòng sử dụng công cụ lựa chọn mã thiết bị để tiếp tục.
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Bạn muốn có nhiều lựa chọn hơn? Vui lòng sử dụng công cụ lựa chọn mã thiết bị để tiếp tục.
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
- Bản vẽ CAD Bản vẽ CAD
- Bản vẽ CAD
-
Giá cho mỗi chi tiết
-
<%range%><%price%>
-
- Các add-on tùy chọn
Công cụ lựa chọn mã thiết bị
Thường mua kèm theo
Phụ kiện