Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.

Spectrometer Memosens Wave CAS80E

Memosens Wave CAS80E: Digital water and wastewater spectrometer for reliable measured values ©Endress+Hauser
Memosens Wave CAS80E: Digital water and wastewater spectrometer with Liquiline CM44 transmitter ©Endress+Hauser

UV-VIS spectrometer for the water & wastewater industry

Memosens Wave CAS80E offers reliable real-time measurements of relevant analytical parameters such as chemical and biological oxygen demand (COD, BOD), turbidity, nitrate and spectral absorption coefficient (SAC) in one single device. It thus increases your efficiency and minimizes costs for your measuring points. The compact, handy Memosens Wave CAS80E spectrometer is easy to install and maintain and ensures reliable, interference-free communication thanks to Memosens digital technology.

  • Simple products

  • Easy to select, install and operate

Technical excellence

Simplicity

  • Standard products

  • Reliable, robust and low-maintenance

Technical excellence

Simplicity

  • High-end products

  • Highly functional and convenient

Technical excellence

Simplicity

  • Specialized products

  • Designed for demanding applications

Technical excellence

Simplicity

Variable

FLEX selections Technical excellence Simplicity
  • F
  • L
  • E
  • X

Fundamental selection

Meet your basic measurement needs

Technical excellence
Simplicity
  • F
  • L
  • E
  • X

Lean selection

Handle your core processes easily

Technical excellence
Simplicity
  • F
  • L
  • E
  • X

Extended selection

Optimize your processes with innovative technologies

Technical excellence
Simplicity
  • F
  • L
  • E
  • X

Xpert selection

Master your most challenging applications

Technical excellence
Simplicity

Variable

  • Specs at a glance

    • Measurement range

      "Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l
      CODeq: 0 to 20 000 mg/l
      BODeq: 0 to 5000 mg/l
      SAC254: 0 to 1000 /m
      TU: 0 to 800 FAU
      TSS: 0 to 10 000 mg/l
      NO3-N: 0 to 500 mg/l
      APHA/Hazen: 0 to 500 Hazen

    • Process temperature

      0 to 50°C
      32 to 122 °F

    • Process pressure

      0.5 to 10 bar abs.
      7.3 to 145 psi

    • Measuring method

      UV/VIS Absorbance at 200 to 800 nm

please wait Loading alternative products

Quick selection

Choose from our suggested options
Sales price
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

or

Individual configuration

Continue your selection from all possible options
  • Benefits

    • Ensures uninterrupted measurements in real time by operating directly in the process.

    • Benefit from reliable, interference-free communication thanks to Memosens digital technology.

    • Offers you the analysis of many standard parameters for your water and...

  • Field of application

    The Memosens Wave CAS80E spectrometer is suitable for measuring a wide range of analytical parameters in:

    • Drinking water

    • Surface water

    • Wastewater

    • Industrial wastewater

    • Utilities

Features and specifications

Back

Details

  • Measuring principle

    Absorption

  • Application

    Monitoring of WWTP in- and outlet

    Drinking water and surface water

  • Installation

    Insitu (immersion in process), extractive with sample and flow cell

  • Characteristic

    Insitu (immersion in process), extractive with sample and flow cell

    UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color

  • Measurement range

    "Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l

    CODeq: 0 to 20 000 mg/l

    BODeq: 0 to 5000 mg/l
    SAC254: 0 to 1000 /m

    TU: 0 to 800 FAU

    TSS: 0 to 10 000 mg/l
    NO3-N: 0 to 500 mg/l
    APHA/Hazen: 0 to 500 Hazen

  • Measuring principle

    UV/VIS absorbance at 200 to 800 nm

  • Design

    40 mm

  • Material

    Housing: stainless steel or titanium

    Windows: sapphire or quartz

    O-rings: EPDM

  • Dimension

    571.4 x 40 mm

    21.5 x 1.6 in

  • Process temperature

    0 to 50°C

    32 to 122 °F

  • Process pressure

    0.5 to 10 bar abs.

    7.3 to 145 psi

  • Temperature sensor

    -20 to 60°C

    -4 to 140 °F

  • Connection

    G1 and NPT 3/4"

  • Ingres protection

    IP68

  • Measuring principle

    UV absorbance

  • Characteristic

    Insitu (immersion in process), extractive with sample and flow cell UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color

  • Measuring method

    UV/VIS Absorbance at 200 to 800 nm

  • Size

    571.4 x 40 mm

    21.5 x 1.6 in

  • Design

    40 mm

  • Process temperature

    0 to 50°C

    32 to 122 °F

  • Ambient temperature

    -20 to 60°C

    -4 to 140 °F

  • Process pressure

    0.5 to 10 bar abs.

    7.3 to 145 psi

  • Specials

    G1, NPT3/4 IP68 (1m/3,3ft) water column with 1N KCl during 60 days

  • Application

    Monitoring of WWTP in- and outlet

    Drinking water and surface water

  • Measurement range

    Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l

    CODeq: 0 to 20 000 mg/l

    BODeq: 0 to 5000 mg/l
    SAC254: 0 to 1000 /m

    TU: 0 to 800 FAU

    TSS: 0 to 10 000 mg/l
    NO3-N: 0 to 500 mg/l
    APHA/Hazen: 0 to 500 Hazen

  • Measuring principle

    UV absorbance

  • Application

    Monitoring of WWTP in- and outlet, drinking water and surface water

  • Characteristic

    Insitu (immersion in process)

    Extractive with sample and flow cell UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color

  • Measurement range

    Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l

    CODeq: 0 to 20 000 mg/l

    BODeq: 0 to 5000 mg/l
    SAC254: 0 to 1000 /m

    TU: 0 to 800 FAU

    TSS: 0 to 10 000 mg/l
    NO3-N: 0 to 500 mg/l
    APHA/Hazen: 0 to 500 Hazen

  • Design

    40 mm

  • Material

    Housing: stainless steel or titanium

    Windows: sapphire or quartz

    O-rings: EPDM

  • Process temperature

    0 to 50°C

    32 to 122 °F

  • Process pressure

    0.5 to 10 bar abs.

    7.3 to 145 psi

  • Temperature sensor

    -20 to 60°C

    -4 to 140 °F

  • Connection

    G1 and NPT 3/4"

  • Ingres protection

    IP68

Documents / Manuals / Software

Accessories / Spare parts

Nr
Order code
Specification

Please select your market / region