Tổng quan về thông số kỹ thuật
-
Measurement range
Ammonium:
0.1 to 1000 mg/l (NH4-N)
Nitrate:
0.1 to 1000 mg/l (NO3-N)
Potassium:
1 to 1000 mg/l
Chloride:
1 to 1000 mg/l -
Process temperature
2 °C to 40°C (36 to 100 °F)
-
Process pressure
400 mbar (160 in H2O) max. permitted overpressure
Phạm vi ứng dụng
ISEmax CAS40D is the right technology for trend measurement directly in the basin. It helps you to increase the efficiency and reduce the energy costs of your aeration processes. The sensor works without reagents saving you even further money. With its robust ion-selective membranes and excellent self-cleaning, it eases your maintenance tasks. Thanks to Memosens, CAS40D combines maximum process and data integrity with best operating comfort and simplifies predictive maintenance.
ISEmax CAS40D measures ammonium, nitrate and pH in municipal wastewater treatment plants:
Aeration basins
- Determination of ammonium and nitrate concentration
- Control and regulation of ammonium decomposition
- Load-dependent aeration control for optimization of the treatment processInlet of aeration basins
- Determination of ammonium load (pH-compensated)
Lợi ích
Nitrate and ammonium measured directly without the need for expensive sample conditioning
Optional potassium and/or chloride measurement compensates for cross sensitivities and ensures reliable measuring results
Low maintenance thanks to automatic flushing unit
Installed directly on the basin rim, no measuring container or sample-conveying pump required
Easy handling due to storage of sensor-specific data
Lựa chọn Extended
-
Sản phẩm cao cấp
- Chức năng cao và tiện lợi
Công nghệ vượt trội
Đơn giản

-
Sản phẩm chuyên dụng
- Dành cho các ứng dụng đòi hỏi cao
Công nghệ vượt trội
Đơn giản
Có thể thay đổi
Lựa chọn FLEX | Công nghệ vượt trội | Đơn giản |
---|---|---|
Lựa chọn Fundamental Đáp ứng nhu cầu đo lường cơ bản |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Lean Xử lý dễ dàng các quy trình cốt lõi |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Extended Tối ưu hóa quy trình bằng công nghệ tiên tiến |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Xpert Làm chủ các ứng dụng khó khăn nhất |
Công nghệ vượt trội
|
Đơn giản
Có thể thay đổi |
Sản phẩm cũ
-
Measuring principle
Potentiometric / ISE
-
Characteristic
Online measurement of nutrient parameters in
municipal wastewater treatment plants
Ion-selective electrode system for the continuous measurement of
ammonium and nitrate -
Size
Diameter: 112 mm (4.41 inch)
Immersion depth: 510mm (20.08 inch) -
Design
Ion-selective electrode(s) for ammonium and/or nitrate, potassium or chloride
pH - (Reference) electrode
Ion-selective electrode for compensating cross-interference
Temperature sensor -
Process temperature
2 °C to 40°C (36 to 100 °F)
-
Ambient temperature
-20 °C to 50 °C (-4 to 120 °F)
-
Process pressure
400 mbar (160 in H2O) max. permitted overpressure
-
Application
Wastewater treatment plants
-
Measurement range
Ammonium:
0.1 to 1000 mg/l (NH4-N)
Nitrate:
0.1 to 1000 mg/l (NO3-N)
Potassium:
1 to 1000 mg/l
Chloride:
1 to 1000 mg/l
Định cấu hình để tải xuống các bản vẽ CAD của sản phẩm này
Chọn theo đề xuất của chúng tôi
-
-- selected coding-- selected label99.999,00<%label%>
-
Bạn muốn có nhiều lựa chọn hơn? Vui lòng sử dụng công cụ lựa chọn mã thiết bị để tiếp tục.
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Bạn muốn có nhiều lựa chọn hơn? Vui lòng sử dụng công cụ lựa chọn mã thiết bị để tiếp tục.
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
-
Giá cho mỗi chi tiết
-
<%range%><%price%>
-
- Các add-on tùy chọn
Công cụ lựa chọn mã thiết bị
Thường mua kèm theo
Phụ kiện
Sản phẩm liên quan