Sản phẩm không có tại thị trường này.
Tổng quan về thông số kỹ thuật
-
Measured Variables
Concentration
Cell Pressure
Cell Temperature -
Hazardous area approvals
ATEX Zone 2
Phạm vi ứng dụng
The SS2100a gas analyzer measures accurate trace gas components (H2O, CO2, H2S, NH3, and C2H2) in natural gas and hydrocarbon processing applications using tunable diode laser absorption spectroscopy (TDLAS) technology. It requires little maintenance and does not need recalibration or periodic replacement parts. The SS2100a is certified for ATEX Zone 2.
Gas analysis: ppb level measurement of H2O & NH3, ppm level measurement of H2O, H2S, CO2, NH3, and C2H2, and percent (%) level measurement of H2S and CO2
Natural Gas: production, storage, transportation, distribution
Refinery: hydrogen recycle, catalytic reforming, flaring, cracking
Gas Processing: fractionation, sour/acid gas removal
LNG: liquefication, custody transfer
Petrochemical: cracking, acetylene converter, caustic wash tower
Lợi ích
Accurate, real-time measurements
Low cost of ownership; no consumables; virtually maintenance free and reliable in harsh environments
Measurements not hampered by wet-up (absorption) or dry-down (desorption)
Dependable with greater repeatability than surface-based sensors
Analog and serial outputs for remote monitoring
Available for the following measurements: H2O, CO2, H2S, NH3, C2H2
ATEX Zone 2 certified
Lựa chọn Xpert
-
Sản phẩm chuyên dụng
- Dành cho các ứng dụng đòi hỏi cao
Công nghệ vượt trội
Đơn giản
Có thể thay đổi

-
Sản phẩm chuyên dụng
- Dành cho các ứng dụng đòi hỏi cao
Công nghệ vượt trội
Đơn giản
Có thể thay đổi
Lựa chọn FLEX | Công nghệ vượt trội | Đơn giản |
---|---|---|
Lựa chọn Fundamental Đáp ứng nhu cầu đo lường cơ bản |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Lean Xử lý dễ dàng các quy trình cốt lõi |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Extended Tối ưu hóa quy trình bằng công nghệ tiên tiến |
Công nghệ vượt trội
|
Đơn giản
|
Lựa chọn Xpert Làm chủ các ứng dụng khó khăn nhất |
Công nghệ vượt trội
|
Đơn giản
Có thể thay đổi |
-
Measuring principle
TDLAS
-
Product Headline
An advanced gas analyzer for hydrogen sulfide (H2S) measurement. The SS2100a uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
-
Channels
1
-
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0-10 to 0-1000 ppmv; 0-5000 ppmv to 0-5%
-
Measured Variables
Concentration
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 50°C (-4 to 122°F)
Optional: -10 to 60°C (14 to 140°F) -
Operating Pressure range
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional) -
Analyzer Wetted materials
316L Stainless Steel
FKM O-Rings
Glass -
Power supply
120 or 240 VAC ±10%, 50-60 Hz; 60W max (with 2 solenoids) - electronics enclosure
120 or 240 VAC, 50-60 Hz - standard; 100W or 200W max for heated systems - sample cabinet -
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms @ 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request -
Housing materials
Electronics: Copper-free Aluminum
Sample System Enclosure: 304 or 316 Stainless Steel -
Hazardous area approvals
ATEX Zone 2
-
Degree of protection
IP66
-
Product safety
CE
-
Measuring principle
TDLAS
-
Product Headline
An advanced gas analyzer for moisture (H2O) measurement. The SS2100a uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
-
Channels
1
-
Analyte and Measurement ranges
H2O (Moisture): 0-10 to 0-100 ppmv; 0-50 to 0-5000 ppmv
-
Measured Variables
Concentration
Moisture Dew Point
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 50°C (-4 to 122°F)
Optional: -10 to 60°C (14 to 140°F) -
Operating Pressure range
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional) -
Analyzer Wetted materials
316L Stainless Steel
FKM O-Rings
Glass -
Power supply
120 or 240 VAC ±10%, 50-60 Hz; 60W max (with 2 solenoids) - electronics enclosure
120 or 240 VAC, 50-60 Hz - standard; 100W or 200W max for heated systems - sample cabinet -
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms @ 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request -
Housing materials
Electronics: Copper-free Aluminum
Sample System Enclosure: 304 or 316 Stainless Steel -
Hazardous area approvals
ATEX Zone 2
-
Degree of protection
IP66
-
Product safety
CE
-
Measuring principle
TDLAS
-
Product Headline
An advanced gas analyzer for carbon dioxide (CO2) measurement. The SS2100a uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
-
Channels
1
-
Analyte and Measurement ranges
CO2 (Carbon Dioxide): 0-10 to 0-1000 ppmv; 0-5000 ppmv to 0-5%
-
Measured Variables
Concentration
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 50°C (-4 to 122°F)
Optional: -10 to 60°C (14 to 140°F) -
Operating Pressure range
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional) -
Analyzer Wetted materials
316L Stainless Steel
FKM O-Rings
Glass -
Power supply
120 or 240 VAC ±10%, 50-60 Hz; 60W max (with 2 solenoids) - electronics enclosure
120 or 240 VAC, 50-60 Hz - standard; 100W or 200W max for heated systems - sample cabinet -
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms @ 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request -
Housing materials
Electronics: Copper-free Aluminum
Sample System Enclosure: 304 or 316 Stainless Steel -
Hazardous area approvals
ATEX Zone 2
-
Degree of protection
IP66
-
Product safety
CE
-
Measuring principle
TDLAS
-
Product Headline
An advanced gas analyzer for ammonia (NH3) measurement. The SS2100a uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
-
Channels
1
-
Analyte and Measurement ranges
NH3 (Ammonia): 0-5 ppmv
-
Measured Variables
Concentration
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 50°C (-4 to 122°F)
Optional: -10 to 60°C (14 to 140°F) -
Operating Pressure range
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional) -
Analyzer Wetted materials
316L Stainless Steel
FKM O-Rings
Glass -
Power supply
120 or 240 VAC ±10%, 50-60 Hz; 60W max (with 2 solenoids) - electronics enclosure
120 or 240 VAC, 50-60 Hz - standard; 100W or 200W max for heated systems - sample cabinet -
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms @ 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request -
Housing materials
Electronics: Copper-free Aluminum
Sample System Enclosure: 304 or 316 Stainless Steel -
Hazardous area approvals
ATEX Zone 2
-
Degree of protection
IP66
-
Product safety
CE
-
Measuring principle
TDLAS
-
Product Headline
An advanced gas analyzer for acetylene (C2H2) measurement. The SS2100a uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
-
Channels
1
-
Analyte and Measurement ranges
C2H2 (Acetylene): 0-5; 0-3000 ppmv
-
Measured Variables
Concentration
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 50°C (-4 to 122°F)
Optional: -10 to 60°C (14 to 140°F) -
Operating Pressure range
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional) -
Analyzer Wetted materials
316L Stainless Steel
FKM O-Rings
Glass -
Power supply
120 or 240 VAC ±10%, 50-60 Hz; 60W max (with 2 solenoids) - electronics enclosure
120 or 240 VAC, 50-60 Hz - standard; 100W or 200W max for heated systems - sample cabinet -
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms @ 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request -
Housing materials
Electronics: Copper-free Aluminum
Sample System Enclosure: 304 or 316 Stainless Steel -
Hazardous area approvals
ATEX Zone 2
-
Degree of protection
IP66
-
Product safety
CE
Định cấu hình để tải xuống các bản vẽ CAD của sản phẩm này
Chọn theo đề xuất của chúng tôi
-
-- selected coding-- selected label99.999,00<%label%>
-
Bạn muốn có nhiều lựa chọn hơn? Vui lòng sử dụng công cụ lựa chọn mã thiết bị để tiếp tục.
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Bạn muốn có nhiều lựa chọn hơn? Vui lòng sử dụng công cụ lựa chọn mã thiết bị để tiếp tục.
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
-
Giá cho mỗi chi tiết
-
<%range%><%price%>
-
- Các add-on tùy chọn